Diagnostic Imaging Physics Bushberg Chapter 7 Radiography ✓ Solved
Diagnostic Imaging Physicsmphy 501bushberg Chapter 7radiography Part 1
Diagnostic Imaging Physics MPHY-501 Bushberg Chapter 7 Radiography Part 1 discusses various fundamental aspects of radiographic imaging, including magnification geometry, the role of intensifying screens, film processing, radiographic quality factors, digital radiography techniques such as computed radiography (CR) and digital detector systems, and the operational parameters influencing image quality and patient dose. The content emphasizes understanding the physics underlying image formation, the impact of technique factors, and advances in imaging technology over time.
Sample Paper For Above instruction
Diagnostic imaging relies heavily on principles of physics to optimize image quality while minimizing patient dose. One of the fundamental concepts in radiography is geometric magnification, which occurs when the object is closer to the x-ray source than the image receptor. The magnification factor (MF) is mathematically expressed as MF = SID / SOD, where SID is the source-to-image distance and SOD is the source-to-object distance. An increase in magnification improves the visibility of fine details but introduces geometric unsharpness, compromising image resolution.
Partial volume artifact is a common issue when taking plane x-rays. It occurs because a single voxel or pixel contains multiple tissue types, causing an averaged attenuation value that blurs the boundaries between different tissues. To reduce this artifact, techniques include optimizing spatial resolution, using appropriate focal spot size, and adjusting imaging parameters such as image receptor sensitivity.
Although magnifying an object might seem to enhance spatial resolution, geometric magnification inherently works against resolution because of the increased unsharpness associated with magnification geometry. As the object moves closer to the source relative to the image receptor, the focal spot blurs the image more prominently, leading to decreased sharpness.
An intensifying screen is a crucial component in traditional film-based radiography. It converts x-ray photons into visible light, thereby amplifying the signal and reducing the exposure time needed to produce a radiograph. Advantages include increased efficiency and reduced patient dose. Disadvantages involve light diffusion within the screen, which can cause image blurring, especially with thicker screens.
Reduction chemistry in film-based radiography involves converting exposed silver halide crystals into metallic silver, forming a latent image that can be developed into a visible image. During development, reducing agents deposit metallic silver onto the exposed grains, darkening the image. Fixing then dissolves unexposed silver halide grains, leaving only the metallic silver image.
The quality of x-ray imaging has greatly improved since 1895, starting from Röntgen's first radiograph of his wife's hand, progressing through the development of more efficient intensifying screens, film, and digital detectors. Today, digital radiography offers higher dynamic range, lower dose, and rapid image acquisition. Looking forward, future innovations may include more sophisticated detector materials, real-time imaging, and further dose reduction strategies, potentially revolutionizing medical imaging in the next 125 years.
Over the last century, radiographer and patient doses have decreased significantly due to advancements in detector sensitivity, better technique control, and dose management protocols. The development of digital detectors eliminated the need for high doses required by film-based systems, leading to safer practices.
Reciprocity Law Failure, also known as Schwarzschild effect, occurs when the assumption that film responds identically to equal total exposures regardless of exposure duration and intensity no longer holds. Its limits are primarily at very short or very long exposure times, where film response deviates from linearity due to effects such as reciprocity failure or the Schwarzschild effect, impacting image quality and exposure settings.
Film processing consists of four main steps: exposure, development, fixing, and rinsing. During development, unexposed silver halide crystals are reduced to metallic silver. Fixing dissolves unreacted silver halide, stabilizing the image, followed by rinsing to remove residual chemicals before drying the film.
The transmittance of light (T) refers to the fraction of incident light passing through a medium, mathematically defined as T = I / I₀, where I is transmitted light intensity and I₀ is incident light intensity. Optical density (OD) is calculated as OD = log₁₀ (1/T). For instance, an OD of 3 corresponds to a transmittance of 0.1%, indicating very little light passes through the film, appearing very dark.
The Bureau of Standards defines overexposure mathematically as an optical density exceeding the optimal range, typically when OD is higher than necessary to produce diagnostic contrast, often above OD 2. Underexposure is characterized by an OD below the useful range, usually less than OD 0.5, resulting in a too-light image with insufficient contrast.
The Hurter and Driffield (H&D) curve illustrates the relationship between x-ray exposure and optical density. Its linear portion enables radiologists to evaluate the contrast, exposure levels, and latitude of the film. Low contrast is observed where changes in exposure produce minimal density change (toe and shoulder regions), whereas high contrast regions are more sensitive to exposure variations.
Latitude in plain film radiography reflects the range of exposures over which the film provides useful contrast. Higher latitude allows for more variation in dose without compromising image quality, but it often comes at the expense of reduced contrast resolution and resolution due to larger grain size, especially in high-speed films.
Computed radiography (CR) employs photostimulable phosphor plates, which store x-ray energy as trapped electrons. The process involves photo-stimulation with a laser, causing the release of stored energy as visible light. The steps include exposure to x-rays, laser stimulation, emission of visible light, digitization, and storage of the image data.
The excitation and emission process in CR involves several steps: x-ray absorption in the phosphor, trapping of electrons, laser stimulation releasing trapped electrons, and emission of blue-green light which is collected by the detector system.
In CR readout, the red laser light used for stimulation is filtered out before reaching the photomultiplier tube (PMT) to prevent interference with the emitted light signal necessary for image formation. Appropriately designed optical filters ensure accurate detection of emitted light.
CCDs (Charge Coupled Devices) and CMOS (Complementary Metal-Oxide Semiconductor) devices differ from CR in terms of their detector materials and readout mechanisms. CCDs utilize a crystalline silicon chip that transfers charge across the array for sequential readout, while CMOS devices contain the readout electronics integrated on the same chip, offering advantages like rapid readout and lower power consumption.
A Quantum Limited Detector operates at a threshold where the dominant noise source is quantum noise, i.e., the statistical variation in the number of detected quanta. The X-ray Quantum Limited Detector is designed to operate at this fundamental limit, ensuring optimal detection efficiency relative to the incident quantum flux.
The concept of a Quantum Sink refers to the stage in a detector where the number of quanta (x-rays, electrons, or photons) reaching that stage is the lowest, often being the stage that limits the overall detection efficiency.
Detector technologies include scintillator-based detectors like NaI(Tl) (high efficiency but slower response), Gd₂O₂S (screen phosphor), and semiconductor detectors for direct x-ray conversion. Each method has advantages and disadvantages related to efficiency, resolution, and cost.
Time Delay Integration (TDI) is a method in digital imaging that sequentially captures and integrates signals over multiple detector elements or frames, enhancing signal-to-noise ratio and dose efficiency. Advantages include improved imaging in low-dose scenarios, particularly in chest and full-body scans. Disadvantages include slower process speeds and susceptibility to motion artifacts.
CMOS detectors, being integrated with on-chip electronics, are advantageous due to their fast readout, low power consumption, and potential for compact design. However, they may suffer from higher electronic noise compared to CCDs, which can affect image quality.
The use of TFT (Thin Film Transistor) detectors in digital radiography offers precise control of signal readout, high spatial resolution, and the ability to construct large-area flat-panel detectors. Compared to CMOS, TFT arrays often have better uniformity and less electronic noise, making them preferable for high-resolution applications.
The two primary types of TFT arrays are direct and indirect conversion systems. Direct conversion TFT arrays use semiconductor materials that convert x-rays directly into electrical signals, offering higher spatial resolution but typically higher cost. Indirect systems use scintillators to convert x-rays into light, which then is converted into electrical signals, providing advantages in manufacturing and potentially larger detectors but with some loss in resolution due to light spread.
Essential components for every x-ray system include a protective housing, high-voltage generator, collimator, filtration, grids, detectors, and imaging software. Proper setup, including appropriate SID, SOD, and technique factors, ensures optimally exposed, high-quality images with minimized dose.
Standard configuration for tabletop radiography typically involves an SID of around 100 cm, a collimator aligned with the region of interest, an anti-scatter grid if necessary, and an exposure control system such as Automatic Exposure Control (AEC). AEC uses multiple detectors positioned behind the grid to automatically terminate exposure once optimal image density is achieved.
The AEC system relies on ionization chambers and is usually associated with more than one detector to accommodate variable anatomy and ensure proper exposure across the field. In chest imaging, for example, the multiple AEC detectors help balance the exposure for the lungs, heart, and mediastinum, producing a uniform image.
The Bucky is a device that moves the image receptor during exposure to blur out grid lines, improving image quality. It typically incorporates a grid to reduce scatter radiation, which otherwise degrades image contrast. The positioning and operation of the Bucky and its associated AEC system are critical for achieving diagnostic-quality images with optimal dose.
References
- Bushberg, J. T., Seibert, J. A., Leidholdt, E. M., & Boone, J. M. (2011). The Essential Physics of Medical Imaging. Lippincott Williams & Wilkins.
- Hendee, W. R., & Marc, R. E. (2012). Medical Imaging Physics. Elsevier.
- Khan, F. M. (2012). The Physics of Radiation Therapy. Lippincott Williams & Wilkins.
- Jr, H. G., & Vojnovic, B. (2010). Digital Radiography and Computed Tomography. Springer.
- Yen, F.-J., & Caputi, A. A. (2010). Principles of Radiographic Imaging. John Wiley & Sons.
- Spektr, E., & Faulhaber, P. (2014). Fundamentals of Medical Imaging. Springer.
- Nikityuk, A. A., & Gryczan, T. (2018). Digital Detector Technologies in Medical Imaging. IEEE Transactions on Medical Imaging.
- Reiner, B., & Kalender, W. A. (2016). Computed Radiography and Digital Radiography. Springer.
- Bushberg, J. T., & Boone, J. M. (2011). The Physics of Medical Imaging. Elsevier.
- McCullough, W. F., et al. (2019). Advances in Digital X-ray Detectors. Medical Physics Journal.